1. After using the oil lens, the residual lens oil on the oil lens should be cleaned, the biological microscope', and a dust cover should be added after the upright biological microscope is used; 2. Frequently use a brush or a damp cloth with neutral detergent to clean or wipe the microscope Non-optical surface, keep the surface clean; 3. When disassembling and assembling the microscope components, please pay attention to avoid touching the optical surface with your fingers; 4. Do not put the eyepiece upside down to avoid dust contamination of the eyepiece lens; 5. Do not direct sunlight , High temperature or high humidity, dusty, and places subject to strong vibration; 6. When moving the microscope, you must carefully hold the handle behind the mirror arm and the mirror base with both hands, and keep it vertical; 7. Halogen lamp 1. When switching the power supply of the tungsten filament light source, the brightness should be adjusted to the minimum; 8. Once the mercury lamp light source is started, it cannot be turned off within less than 30 minutes. After turning off the high-pressure mercury lamp, it can be turned on again at least 10 minutes later.